- Catalog number
- FEK0773
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- Kunitz-type protease inhibitor 2 is an enzyme inhibitor that in humans is encoded by the SPINT2 gene. This gene encodes a transmembrane protein with two extracellular Kunitz domains that inhibits a variety of serine proteases. The protein inhibits HGF activator which prevents the formation of active hepatocyte growth factor. And this gene is a putative tumor suppressor, and mutations in this gene result in congenital sodium diarrhea. Multiple transcript variants encoding different isoforms have been found for this gene.
- Reproducibility
- Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 59 51 45 47 51 2.22 4.4% Sample 2 211 221 223 199 214 8.33 3.9% Sample 3 923 961 826 917 907 53.5 5.9% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 59 51 45 47 51 2.22 4.4% Sample 1 59 51 45 47 51 2.22 4.4% Sample 2 211 221 223 199 214 8.33 3.9% Sample 2 211 221 223 199 214 8.33 3.9% Sample 3 923 961 826 917 907 53.5 5.9% Sample 3 923 961 826 917 907 53.5 5.9%
- Gene Name
- CST3
- Gene ID
- 10653
- UniProt
- O43291
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: NS0; Immunogen sequence: M1-K197
- Capture Antibody
- Polyclonal antibody from goat
- Detection Antibody
- Polyclonal antibody from goat
- Applications
- ELISA
- Field of Research
- Cardiovascular, Caspases, Cell Biology, Cell Type Markers, Neural Stem Cells, Stem Cells, Tags & Cell Markers, Tumor Associated
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, serum, plasma (heparin, EDTA) and cell lysates
- Detection Range
- 31.2 pg/mL - 2,000 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 59 211 923 61 224 890 Standard deviation 3.07 10.97 61.84 3.54 12.32 62.3 CV (%) 5.2% 5.2% 6.7% 5.8% 5.5% 7% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 59 211 923 61 224 890 Standard deviation 3.07 10.97 61.84 3.54 12.32 62.3 CV (%) 5.2% 5.2% 6.7% 5.8% 5.5% 7% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 59 211 923 61 224 890 Mean (pg/ml) 59 211 923 61 224 890 Standard deviation 3.07 10.97 61.84 3.54 12.32 62.3 Standard deviation 3.07 10.97 61.84 3.54 12.32 62.3 CV (%) 5.2% 5.2% 6.7% 5.8% 5.5% 7% CV (%) 5.2% 5.2% 6.7% 5.8% 5.5% 7%
- Sensitivity
- <10 pg/mL
- Components
- Description|Quantity|Volume Anti-tag Pre-coated 96-well Strip Microplate|1|12 strips of 8 wells Standard||2 Antibody A|1|3ml Antibody B|1|3ml Sample Diluent|1|15ml TBS-T Wash Buffer (25x) |1|12ml Color Developing Reagent (TMB) |1|10ml Stop Solution|1|10ml Adhesive Plate Sealers|2|Piece
- Storage Conditions
- Store at 4°C for 6 months. (Ships with gel ice, can store for up to 3 days in room temperature. Refrigerate upon receipt.)
- Specificity
- Natural and recombinant Human SPINT2
- Anticoagulant
- Heparin or EDTA
- Gene Name Synonym
- Serine peptidase inhibitor, Kunitz type, 2
- Protein Name
- Kunitz-type protease inhibitor 2
- Sample Data
- Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.096 0.157 0.202 0.298 0.473 0.787 1.322 2.105 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.096 0.157 0.202 0.298 0.473 0.787 1.322 2.105 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.096 0.157 0.202 0.298 0.473 0.787 1.322 2.105 O.D. 0.096 0.157 0.202 0.298 0.473 0.787 1.322 2.105
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Automated plate washer (optional) Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution. Horizontal orbital microplate shaker capable of maintaining a speed of 500 rpm, amplitude 3mm.