- Catalog number
- FEK0739
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- C-X-C motif chemokine 13 (CXCL13), also known as B lymphocyte chemoattractant (BLC), is a protein that in humans is encoded by theCXCL13 gene. CXCL13 is a small cytokine belonging to the CXC chemokine family, and it was located on sequence maps to 4q21, near the genes encoding most CXC chemokines. In T-lymphocytes, CXCL13 expression is thought to reflect a germinal center origin of the T-cell. Hence, expression of CXCL13 in T-cell lymphomas, such as Angioimmunoblastic T-cell Lymphoma, is thought to reflect a germinal center origin of the neoplastic T-cells
- Reproducibility
- Reproducibility To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots. Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 36 41 38 34 37 2.57 6.9% Sample 2 186 197 198 189 193 5.2 2.7% Sample 3 437 495 459 516 477 11.44 2.4% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 36 41 38 34 37 2.57 6.9% Sample 1 36 41 38 34 37 2.57 6.9% Sample 2 186 197 198 189 193 5.2 2.7% Sample 2 186 197 198 189 193 5.2 2.7% Sample 3 437 495 459 516 477 11.44 2.4% Sample 3 437 495 459 516 477 11.44 2.4%
- Gene Name
- CXCL13
- Gene ID
- 10563
- UniProt
- O43927
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: E.coli; Immunogen sequence: V23-R94
- Capture Antibody
- Polyclonal antibody from goat
- Detection Antibody
- Polyclonal antibody from goat
- Applications
- ELISA
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, serum, plasma (heparin, EDTA) and saliva
- Detection Range
- 15.6 pg/mL - 1,000 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 36 186 437 34 191 423 Standard deviation 2.19 9.67 28.84 2.1 12.6 33.41 CV (%) 6.1% 5.2% 6.6% 6.2% 6.6% 7.9% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 36 186 437 34 191 423 Standard deviation 2.19 9.67 28.84 2.1 12.6 33.41 CV (%) 6.1% 5.2% 6.6% 6.2% 6.6% 7.9% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 36 186 437 34 191 423 Mean (pg/ml) 36 186 437 34 191 423 Standard deviation 2.19 9.67 28.84 2.1 12.6 33.41 Standard deviation 2.19 9.67 28.84 2.1 12.6 33.41 CV (%) 6.1% 5.2% 6.6% 6.2% 6.6% 7.9% CV (%) 6.1% 5.2% 6.6% 6.2% 6.6% 7.9%
- Sensitivity
- <1 pg/mL
- Components
- Description|Quantity|Volume Anti-tag Pre-coated 96-well Strip Microplate|1|12 strips of 8 wells Standard||2 Antibody A|1|3ml Antibody B|1|3ml Sample Diluent|1|15ml TBS-T Wash Buffer (25x) |1|12ml Color Developing Reagent (TMB) |1|10ml Stop Solution|1|10ml Adhesive Plate Sealers|2|Piece
- Storage Conditions
- Store at 4°C for 6 months. (Ships with gel ice, can store for up to 3 days in room temperature. Refrigerate upon receipt.)
- Specificity
- Natural and recombinant Human CXCL13
- Anticoagulant
- Heparin or EDTA
- Sample Data
- Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1,000 O.D. 0.052 0.140 0.187 0.311 0.516 0.916 1.448 2.196 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1,000 O.D. 0.052 0.140 0.187 0.311 0.516 0.916 1.448 2.196 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1,000 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1,000 O.D. 0.052 0.140 0.187 0.311 0.516 0.916 1.448 2.196 O.D. 0.052 0.140 0.187 0.311 0.516 0.916 1.448 2.196
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Automated plate washer (optional) Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution. Horizontal orbital microplate shaker capable of maintaining a speed of 500 rpm, amplitude 3mm.