- Catalog number
- FEK0809
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- Chemokine (C-C motif) ligand 13 (CCL13) is a small cytokine belonging to the CC chemokine family. Its gene is located on human chromosome 17 within a large cluster of other CC chemokines. CCL13 induces chemotaxis in monocytes, eosinophils, T lymphocytes, and basophils by binding cell surface G-protein linked chemokine receptors such as CCR2, CCR3 and CCR5. Activity of this chemokine has been implicated in allergic reactions such as asthma. CCL13 can be induced by the inflammatory cytokines interleukin-1 and TNF-alpha.
- Reproducibility
- Reproducibility To assay reproducibility, three samples with differing target protein concentrations were assayed using four different lots. Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 19 18 17 17 18 1.26 7.1% Sample 2 87 85 88 82 86 2.05 2.4% Sample 3 258 234 248 251 248 11.15 4.5% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 19 18 17 17 18 1.26 7.1% Sample 1 19 18 17 17 18 1.26 7.1% Sample 2 87 85 88 82 86 2.05 2.4% Sample 2 87 85 88 82 86 2.05 2.4% Sample 3 258 234 248 251 248 11.15 4.5% Sample 3 258 234 248 251 248 11.15 4.5%
- Gene Name
- CCL13
- Gene ID
- 6357
- UniProt
- Q99616
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: E.coli; Immunogen sequence: Q24-T98
- Capture Antibody
- Polyclonal antibody from goat
- Detection Antibody
- Polyclonal antibody from goat
- Applications
- ELISA
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, cell lysates, serum and plasma (heparin, EDTA)
- Detection Range
- 7.8 pg/mL - 500 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 19 87 258 18 85 271 Standard deviation 1.16 5.39 17.8 1.26 5.44 15.91 CV (%) 6.1% 6.2% 6.9% 7% 6.4% 7.2% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 19 87 258 18 85 271 Standard deviation 1.16 5.39 17.8 1.26 5.44 15.91 CV (%) 6.1% 6.2% 6.9% 7% 6.4% 7.2% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 19 87 258 18 85 271 Mean (pg/ml) 19 87 258 18 85 271 Standard deviation 1.16 5.39 17.8 1.26 5.44 15.91 Standard deviation 1.16 5.39 17.8 1.26 5.44 15.91 CV (%) 6.1% 6.2% 6.9% 7% 6.4% 7.2% CV (%) 6.1% 6.2% 6.9% 7% 6.4% 7.2%
- Sensitivity
- <10 pg/mL
- Components
- Description|Quantity|Volume Anti-tag Pre-coated 96-well Strip Microplate|1|12 strips of 8 wells Standard||2 Antibody A|1|3ml Antibody B|1|3ml Sample Diluent|1|15ml TBS-T Wash Buffer (25x) |1|12ml Color Developing Reagent (TMB) |1|10ml Stop Solution|1|10ml Adhesive Plate Sealers|2|Piece
- Storage Conditions
- Store at 4°C for 6 months. (Ships with gel ice, can store for up to 3 days in room temperature. Refrigerate upon receipt.)
- Specificity
- Natural and recombinant Human CCL13
- Anticoagulant
- Heparin or EDTA
- Sample Data
- Concentration (pg/ml) 0 7.8 15.6 31.3 62.5 125 250 500 O.D. 0.054 0.198 0.325 0.579 0.992 1.672 2.213 2.565 Concentration (pg/ml) 0 7.8 15.6 31.3 62.5 125 250 500 O.D. 0.054 0.198 0.325 0.579 0.992 1.672 2.213 2.565 Concentration (pg/ml) 0 7.8 15.6 31.3 62.5 125 250 500 Concentration (pg/ml) 0 7.8 15.6 31.3 62.5 125 250 500 O.D. 0.054 0.198 0.325 0.579 0.992 1.672 2.213 2.565 O.D. 0.054 0.198 0.325 0.579 0.992 1.672 2.213 2.565
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Automated plate washer (optional) Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution. Horizontal orbital microplate shaker capable of maintaining a speed of 500 rpm, amplitude 3mm.