- Catalog number
- EK2186
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- Dipeptidyl-peptidase 2 is an enzyme that in humans is encoded by the DPP7 gene. The protein encoded by this gene is a post-proline cleaving aminopeptidase expressed in quiescent lymphocytes. The resting lymphocytes are maintained through suppression of apoptosis, a state which is disrupted by inhibition of this novel serine protease. The enzyme has strong sequence homology with prolylcarboxypeptidase and is active at both acidic and neutral pH.
- Reproducibility
- Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 86 81 85 89 85 2.73 3.2% Sample 2 212 226 208 215 215 9.47 4.4% Sample 3 922 965 942 951 945 38.75 4.1% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 86 81 85 89 85 2.73 3.2% Sample 1 86 81 85 89 85 2.73 3.2% Sample 2 212 226 208 215 215 9.47 4.4% Sample 2 212 226 208 215 215 9.47 4.4% Sample 3 922 965 942 951 945 38.75 4.1% Sample 3 922 965 942 951 945 38.75 4.1%
- Gene Name
- DPP7
- Gene ID
- 29952
- UniProt
- Q9UHL4
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: NS0; Immunogen sequence: G22-L492
- Capture Antibody
- A monoclonal antibody from mouse
- Detection Antibody
- A biotinylated polyclonal antibody from goat
- Applications
- ELISA
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, serum and plasma (heparin, EDTA)
- Detection Range
- 31.2 pg/mL - 2,000 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 86 212 922 95 226 892 Standard deviation 4.39 11.87 56.24 3.71 10.85 37.46 CV (%) 5.1% 5.6% 6.1% 3.9% 4.8% 4.2% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 86 212 922 95 226 892 Standard deviation 4.39 11.87 56.24 3.71 10.85 37.46 CV (%) 5.1% 5.6% 6.1% 3.9% 4.8% 4.2% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 86 212 922 95 226 892 Mean (pg/ml) 86 212 922 95 226 892 Standard deviation 4.39 11.87 56.24 3.71 10.85 37.46 Standard deviation 4.39 11.87 56.24 3.71 10.85 37.46 CV (%) 5.1% 5.6% 6.1% 3.9% 4.8% 4.2% CV (%) 5.1% 5.6% 6.1% 3.9% 4.8% 4.2%
- Sensitivity
- <10 pg/mL
- Components
- Description|Quantity Pre-coated 96-well strip microplate|1 Standard|2 vials Biotinylated antibody (100x) |100ul Avidin-Biotin-Peroxidase Complex (100x) |100ul Sample Diluent|30ml Antibody Diluent|12ml Avidin-Biotin-Peroxidase Diluent|12ml Color Developing Reagent (TMB) |10ml Stop Solution|10ml Wash Buffer (25x) |20ml Adhesive plate sealers|4
- Storage Conditions
- Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
- Specificity
- Natural and recombinant Human DPP7
- Sample Data
- Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.052 0.12 0.178 0.285 0.507 0.902 1.571 2.684 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.052 0.12 0.178 0.285 0.507 0.902 1.571 2.684 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 Concentration (pg/ml) 0 31.2 62.5 125 250 500 1000 2000 O.D. 0.052 0.12 0.178 0.285 0.507 0.902 1.571 2.684 O.D. 0.052 0.12 0.178 0.285 0.507 0.902 1.571 2.684
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer (optional) . Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution.