- Catalog number
- EK2200
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- This gene is a member of the immunoglobulin gene superfamily and encodes a receptor for the Fc region of IgA. The receptor is a transmembrane glycoprotein present on the surface of myeloid lineage cells such as neutrophils, monocytes, macrophages, and eosinophils, where it mediates immunologic responses to pathogens. It interacts with IgA-opsonized targets and triggers several immunologic defense processes, including phagocytosis, antibody-dependent cell-mediated cytotoxicity, and stimulation of the release of inflammatory mediators. Multiple alternatively spliced transcript variants encoding different isoforms have been described for this gene.
- Reproducibility
- Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 23 25 29 27 26 1.69 6.5% Sample 2 67 68 75 88 75 3.13 4.2% Sample 3 267 296 270 311 286 13.16 4.6% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 23 25 29 27 26 1.69 6.5% Sample 1 23 25 29 27 26 1.69 6.5% Sample 2 67 68 75 88 75 3.13 4.2% Sample 2 67 68 75 88 75 3.13 4.2% Sample 3 267 296 270 311 286 13.16 4.6% Sample 3 267 296 270 311 286 13.16 4.6%
- Gene Name
- FCAR
- Gene ID
- 2204
- UniProt
- P24071
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: NS0; Immunogen sequence: Q22-N227
- Capture Antibody
- A monoclonal antibody from mouse
- Detection Antibody
- A biotinylated polyclonal antibody from goat
- Applications
- ELISA
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, serum and plasma (heparin, EDTA)
- Detection Range
- 15.6 pg/mL - 1,000 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 23 67 267 25 70 275 Standard deviation 1.29 3.89 16.55 1.73 4.97 17.88 CV (%) 5.6% 5.8% 6.2% 6.9% 7.1% 6.5% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 23 67 267 25 70 275 Standard deviation 1.29 3.89 16.55 1.73 4.97 17.88 CV (%) 5.6% 5.8% 6.2% 6.9% 7.1% 6.5% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 23 67 267 25 70 275 Mean (pg/ml) 23 67 267 25 70 275 Standard deviation 1.29 3.89 16.55 1.73 4.97 17.88 Standard deviation 1.29 3.89 16.55 1.73 4.97 17.88 CV (%) 5.6% 5.8% 6.2% 6.9% 7.1% 6.5% CV (%) 5.6% 5.8% 6.2% 6.9% 7.1% 6.5%
- Sensitivity
- <1 pg/mL
- Components
- Description|Quantity Pre-coated 96-well strip microplate|1 Standard|2 vials Biotinylated antibody (100x) |100ul Avidin-Biotin-Peroxidase Complex (100x) |100ul Sample Diluent|30ml Antibody Diluent|12ml Avidin-Biotin-Peroxidase Diluent|12ml Color Developing Reagent (TMB) |10ml Stop Solution|10ml Wash Buffer (25x) |20ml Adhesive plate sealers|4
- Storage Conditions
- Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
- Specificity
- Natural and recombinant human FCAR
- Sample Data
- Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1000 O.D. 0.017 0.147 0.242 0.421 0.757 1.187 1.638 2.001 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1000 O.D. 0.017 0.147 0.242 0.421 0.757 1.187 1.638 2.001 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1000 Concentration (pg/ml) 0 15.6 31.2 62.5 125 250 500 1000 O.D. 0.017 0.147 0.242 0.421 0.757 1.187 1.638 2.001 O.D. 0.017 0.147 0.242 0.421 0.757 1.187 1.638 2.001
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer (optional) . Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution.