- Catalog number
- FEK1312
- Size / count
- 96 Well With Removable Strips
- Dry ice
- No
- Hazardous
- No
- Discontinued
- No
- Background
- CD32 is a surface receptor protein and part of a large population of B cell co-receptors, which act to modulate signaling. It has a low-affinity for IgG antibodies and down-regulates antibody production in the presence of IgG. This feedback loops essentially lowers the production of IgG produced by B cells when there is a surplus in the body.
- Reproducibility
- Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 122 125 105 106 114 9.06 7.9% Sample 2 404 361 357 357 369 19.84 5.3% Sample 3 1372 1213 1353 1369 1326 66.06 4.9% Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Lots Lot 1 (pg/ml) Lot 2 (pg/ml) Lot 3 (pg/ml) Lot 4 (pg/ml) Mean (pg/ml) Standard Deviation CV (%) Sample 1 122 125 105 106 114 9.06 7.9% Sample 1 122 125 105 106 114 9.06 7.9% Sample 2 404 361 357 357 369 19.84 5.3% Sample 2 404 361 357 357 369 19.84 5.3% Sample 3 1372 1213 1353 1369 1326 66.06 4.9% Sample 3 1372 1213 1353 1369 1326 66.06 4.9%
- Synonyms
- Low affinity immunoglobulin gamma Fc region receptor II-c; IgG Fc receptor II-c; CDw32; Fc-gamma RII-c; Fc-gamma-RIIc; FcRII-c; CD32; FCGR2C; CD32, FCG2, IGFR2; FCGR2B
- Gene Name
- FCGR2B
- Gene ID
- 2213/9103
- UniProt
- P31994/P31995
- Reactivity
- Human
- Cross Reactivity
- There is no detectable cross-reactivity with other relevant proteins.
- Immunogen
- Expression system for standard: NS0; Immunogen sequence: A46-P217
- Capture Antibody
- Monoclonal antibody from mouse
- Detection Antibody
- Polyclonal antibody from goat
- Applications
- ELISA
- Standard
- 10 ng/Tube
- Sample Type
- Cell culture supernatants, cell lysates, serum and plasma (heparin, EDTA) .
- Detection Range
- 46.9 pg/mL - 3,000 pg/mL
- Precision
- Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 122 404 1372 133 365 1478 Standard deviation 8.05 19.79 97.41 9.7 22.26 130.06 CV (%) 6.6% 4.9% 7.1% 7.3% 6.1% 8.8% Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (pg/ml) 122 404 1372 133 365 1478 Standard deviation 8.05 19.79 97.41 9.7 22.26 130.06 CV (%) 6.6% 4.9% 7.1% 7.3% 6.1% 8.8% Intra-Assay Precision Inter-Assay Precision Intra-Assay Precision Inter-Assay Precision Sample 1 2 3 1 2 3 Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 n 16 16 16 24 24 24 Mean (pg/ml) 122 404 1372 133 365 1478 Mean (pg/ml) 122 404 1372 133 365 1478 Standard deviation 8.05 19.79 97.41 9.7 22.26 130.06 Standard deviation 8.05 19.79 97.41 9.7 22.26 130.06 CV (%) 6.6% 4.9% 7.1% 7.3% 6.1% 8.8% CV (%) 6.6% 4.9% 7.1% 7.3% 6.1% 8.8%
- Sensitivity
- <10 pg/mL
- Components
- Description|Quantity|Volume Anti-tag Pre-coated 96-well Strip Microplate|1|12 strips of 8 wells Standard||2 Antibody A|1|3ml Antibody B|1|3ml Sample Diluent|1|15ml TBS-T Wash Buffer (25x) |1|12ml Color Developing Reagent (TMB) |1|10ml Stop Solution|1|10ml Adhesive Plate Sealers|2|Piece
- Storage Conditions
- Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
- Calculated Molecular Weight
- 35578 MW
- Specificity
- Natural and recombinant human FCGR2B/C
- Anticoagulant
- Heparin or EDTA
- Gene Name Synonym
- Fc fragment of IgG receptor IIb/c
- Protein Name
- Low affinity immunoglobulin gamma Fc region receptor II-b/c
- Sample Data
- Concentration (pg/ml) 0 46.9 93.8 187.5 375 750 1500 3000 O.D. 0.054 0.125 0.199 0.337 0.619 1.143 1.687 2.050 Concentration (pg/ml) 0 46.9 93.8 187.5 375 750 1500 3000 O.D. 0.054 0.125 0.199 0.337 0.619 1.143 1.687 2.050 Concentration (pg/ml) 0 46.9 93.8 187.5 375 750 1500 3000 Concentration (pg/ml) 0 46.9 93.8 187.5 375 750 1500 3000 O.D. 0.054 0.125 0.199 0.337 0.619 1.143 1.687 2.050 O.D. 0.054 0.125 0.199 0.337 0.619 1.143 1.687 2.050
- Material Required But Not Provided
- Microplate Reader capable of reading absorbance at 450nm. Automated plate washer (optional) Pipettes and pipette tips capable of precisely dispensing 0.5 μl through 1 ml volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500ml graduated cylinders. Test tubes for dilution. Horizontal orbital microplate shaker capable of maintaining a speed of 500 rpm, amplitude 3mm.